Manual

41
all measurements and should be reset for different dilutions or
temperatures.
C. When does it make a lot of difference?
accuracy of any TDS conversion. Assume we have industrial process
uncompensated.
The difference in values is 40%.
In spite of such large error, some users will continue to take data in
the “NaCl” mode because their previous data gathering and process
monitoring was done with an older NaCl referenced device.
Selecting the correct Solution Type on the Ultrameter II will allow the
user to attain true TDS readings that correspond to evaporated weight.
the “User” mode.
XXII. TEMPERATURE COMPENSATION (Tempco)
and TDS DERIVATION
The Ultrameter II contains internal algorithms for characteristics of the
TDS conversion values of a unique solution via the keypad.
A. Conductivity Characteristics
When taking conductivity measurements, the Solution Selection
determines the characteristic assumed as the instrument reports what a