cool. Retest.
or
table salt and tap water.
electrode.
56).
It is possible to desensitize the ORP sensor when measuring
high ORP solutions. To check the ORP sensor functioning, take
an ORP reading of Myron L pH/ORP sensor storage solution. If
the reading is outside the range of 350-400 mV, clean ONLY the
platinum ORP electrode with Myron L ORP Conditioner solution-
soaked cotton swab, being careful not to touch the swab to the
glass bulb of the pH sensor.