XVIII. TROUBLESHOOTING CHART
Symptom Possible Cause Corrective Action
No display
Battery weak or not connected. Check connections or replace battery. Ref. Battery Replacement, pg. 50.
pH readings 1. pH calibration needed. Ref. pH Cal.,
pg. 33.
No response to pH changes
internal crack.
Replace pH/ORP sensor. Ref. Replacement pH/ORP Sensor, pg. 56.
pH 7 pH/ORP sensor has lost KCl.
pH readings drift or respond slowly to
FAC
is displayed repeatedly
Unstable Conductivity/TDS/
Resistivity readings
atmospheric contamination.
Unable to calibrate Conductivity/
TDS
Film or deposits on electrodes.
Resistivity
expected
from pH sensor well.
2. Carbon dioxide in test sample.
place.
Alkalinity or Hardness titration with
calibration
stated on bottle
1. Method error.
1. Repeat titration.
4. Use a new pipette.
Low Reading
Slow or no response to ORP changes
Clean ORP electrode with ORP conditioner. Ref. Cleaning Sensors, 2. pH/ORP
Sensor, pg. 50.