Manual
IV.  INITIAL CALIBRATION and FACTORY SETTINGS
CAUTION: The factory settings are not intended to replace calibration 
with reference solution/buffers.
Factory settings do not take into account the current condition of 
the actual sensor installed in your instrument or any environmental 
conditions that might exist in your facility.
• DO NOT use a Digital Dialysate Meter™ to check the pH or 
conductivity of dialysate using factory calibration settings. 
• ALWAYS perform an initial calibration of a Digital Dialysate 
Meter™ with reference solutions/buffers prior to putting it into use 
forthersttime.
The following table lists initial calibration procedures and their location 
in this manual.
PARAMETER PROCEDURE REFERENCE SOLUTION
General instructions (see Calibration, pg. 16)
Conductivity Gain  14.00 mS NaCl
1
 (see Conductivity or 
TDS Calibration, pg. 19)
Resistivity Electronically calibrated with Conductivity Gain
TDS Gain 442™
1
 (see Conductivity or TDS 
Calibration, pg. 19)
3000 PPM
1 
(see Conductivity or TDS 
Calibration, pg. 19)
pH Zero Point 7.0 pH (see pH Zero Calibration 
(D-6), pg. 20)
Gain – Acid 4.0 pH (see pH Gain Calibration 
(D-6), pg. 21)
Gain – Base 10.0 pH (see pH Gain Calibration 
(D-6), pg. 21)
ORP Electronically calibrated with pH Zero Point
Temperature Does not require calibration
1
Make sure the correct solution type is selected and that it matches the type of 
 reference solution being used (see Solution Selection, pg. 15)
Avoid letting solvents come into contact with the instrument’s sensors, 
case or window areas. Shock damage from a fall may cause the 
instrument to fail.
WARNING:  You MUST verify the function and calibration of any Dialysate 
Meter that has been dropped by performing calibration checks (see 
Checking pH/ORP Sensor Performance, pg. 34) of all measurement 
functions prior to putting it back into use.
8










