Manual
    4. CAL/MCLR Key
CAUTION:  Unintentionally  pressing  the  CAL/MCLR  key  while  making 
measurements could  inadvertently change  the calibration  of  the 
instrument by activating either:
• Thecalibrationmode(the“CAL”iconwillappearontheinstrument’s
display), or
• Thesolutionselectionmenu(theword“SEL”willappearonthe
instrument’s display). 
• Ifyoudoaccidentallyactivateeitherofthesefeatures:
• DO NOT press either the UP or DOWN keys.
• IMMEDIATELY Repress the desired parameter function key 
to return the instrument to measurement mode.
A press of  allows you to enter the calibration mode while
measuring  conductivity, TDS  or pH.  Once  in  “CAL”  mode,  a  press  of 
this key accepts the new value. If no more calibration options follow, the 
instrument returns to measuring (ref. Leaving Calibration, pg. 18).
If  is held down for about 3 seconds, “CAL” mode is not entered, 
but “SEL” appears to allow Solution Selection (ref. pg. 15) with the Up 
or Down keys. As in calibration, the CAL key is now an accept key.
While reviewing stored records, the MCLR side of the key is active to 
allow clearing records (ref. Clearing a Record/Memory Clear, pg. 26).
    5. UP or DOWN Keys
While measuring in any parameter, the or keys activate 
the Memory Store and Memory Recall functions. 
While in “CAL” mode, the keys step or scroll the displayed value up or 
down. A single press steps the display and holding either key scrolls the 
value rapidly.
While in Memory Recall, the keys scroll the display up and down through 
the stack of records (ref. Memory Recall, pg. 25).
VI.  AFTER USING the Dialysate Meter
  A. Maintenance of the Conductivity Cell
Rinse out the conductivity cell with clean water. Do not scrub the cell. 
Foroilylms,squirtinafoamingnon-abrasivecleanerandrinse(ref.
Cleaning Sensors, pg. 34). Even if a very active chemical discolors the 
electrodes, this does not affect the accuracy; 
leave it alone
.
11










