Manual
  B. Measuring Resistivity
Resistivity is for low conductivity solutions. In a cell cup the value may 
drift from trace contaminants or absorption from atmospheric gasses, 
someasuringaowingsampleisrequired.
1.  Ensure pH protective cap is secure to avoid contamination.
2.  Hold instrument at 30° angle (cup sloping downward).
3. Letsampleowcontinuouslyintoconductivitycellwithnoaeration.
4.  Press  key; use best reading.
NOTE:  If reading is lower than 10 kilohms display will be dashes: [ - - - - ].  
Use Conductivity instead.
  C. Measuring pH (D-6)
WARNING:  When using a Digital Dialysate Meter™ to test the pH of 
dialysate, ALWAYS perform a calibration check of the pH sensor at 
the beginning and end of each work day or more often if required by 
your clinic’s internal procedures (see pH Calibration, pg. 20). 
1.  Remove protective cap by grasping its sides and pulling up.
2.  Rinse pH/ORP sensor well and conductivity cell 3 times with 
  sample to be measured. Shake out each sample to remove 
  any residual liquid.
3. RellpH/ORPsensorwellandconductivitycellwithsample.
4.  Press  .
5.  Note value displayed.
6.  IMPORTANT:Afteruse,llpH/ORPsensorwellwithMyronLpH
Sensor Storage Solution and replace protective cap. If Myron L 
pH Sensor Storage Solution is unavailable, you can use a strong 
KCl solution (preferred), a saturated solution of NaCl table salt and 
tap water (next best alternative), or a pH 4 buffer (least desirable 
alternative). NEVER use distilled water or plain tap water as this 
will quickly deplete the sensor (ref. Cleaning Sensors, 3. pH/ORP, 
pg. 35). 
Do not allow pH/ORP sensor to dry out.
  D. Measuring ORP (D-6)
1.  Remove protective cap by grasping its sides and pulling up.
2.  Rinse pH/ORP sensor well and conductivity cell 3 times with 
sample to be measured. Shake out each sample to remove 
any residual liquid.
3. RellboththepH/ORPsensorwellandtheconductivitycell
with sample.
13










