Manual

4. Press .
5. Take reading.
NOTE: When ppm free chlorine units are selected (ref. VIII. ORP UNIT
SELECTION), annunciators alert you when the concentration is outside
thespeciedmeasurementrange.“Or” (over range) will display when
the concentration is over the range limit (> 9.9 ppm). “Ur” (under range)
will display when the concentration is below the range limit (< 0.2 ppm).
WARNING: When using a D-6 to verify the complete removal of residual
oxidizing germicides, such as chlorine, ORP mV reading should be
as low as possible. Generally, an ORP reading of < 275 mV is a good
indicator that there is very little residual oxidizer remaining. However,
it is recommended that a target reading matching or close to the ORP
value measured at the output of your RO system be sought.
6. IMPORTANT:Afteruse,llpH/ORPsensorwellwithMyronLpH
Sensor Storage Solution and replace protective cap. If Myron L
pH Sensor Storage Solution is unavailable, you can use a strong
KCl solution (preferred), a saturated solution of NaCl table salt and
tap water (next best alternative), or a pH 4 buffer (least desirable
alternative). NEVER use distilled water or plain tap water as this
will quickly deplete the sensor (ref. Cleaning Sensors, 3. pH/ORP,
pg. 35).
Do not allow pH/ORP sensor to dry out.
VIII. ORP UNIT SELECTION (D-6) mV OR PPM FREE CHLORINE
1. Press .
2. Press and hold
for approximately 3 seconds.
The current preference for
ORP units of measure is
displayed. Factory setting for
this preference is mV (see
Figure 1).
4. Press the o r keys
to toggle between mV
and ppm free chlorine
measurement format. The
setting chosen is displayed.
(ppm shown as example in
Figure 2).
mV
Figure 1
14
PPM
Figure 2