Manual
Function Solution Recalibrate if 
Dialysate Meter 
Reports Below
Recalibrate if 
Dialysate Meter 
Reports Above
Conductivity 
(NaCl)
14.00 mS 13.86 mS 14.14 mS
Conductivity 
(KCl)
7000µS 6930µS 7070µS
TDS 442™  Solution 
(3000 PPM)
3030 ppm 2970 ppm
pH 7.0 Buffer 6.96 pH 7.04 pH
4.0 Buffer 3.96 pH 4.04 pH
10.0 Buffer 9.96 pH 10.04 pH
Once in “CAL” mode, the  key becomes an accept key. At each
point, pressing   accepts the new calibration value and steps you
to the next adjustment (or out of “CAL” mode if there are no more steps).
To bypass a calibration step, simply press  to accept the present 
value as is.
      b. Leaving Calibration
Calibration  is  complete  when  the  “CAL”  icon  goes  out.  Pressing  any 
measurement key cancels changes not yet accepted and exits calibration 
mode.
Leaving pH after the 2nd buffer results in the same gain being entered 
in place of the 3rd buffer.
    2. Calibration Limits
There  are  calibration  limits. A  nominal  “FAC”  value  is  an  ideal  value 
stored  by  the  factory. Attempts  to  calibrate  too  far,  up  or  down,  from 
there will  cause the displayed value to  be replaced with “FAC”. If 
you  accept  it  (press  the  CAL  key),  you  will  have  the  original  default 
factory  calibration  for  this  measurement.  The  need  to  calibrate  out 
so  far  that  “FAC”  appears  indicates  a  procedural  problem,  incorrect 
standard solution, a very dirty cell cup or an aging pH/ORP sensor (ref. 
Troubleshooting Chart, pg. 37-38).
WARNING:
• Factorysettingsdonottakeintoaccountthecurrentconditionof
the sensor installed in your instrument.
• Ifyouareunabletoadjustthecalibrationvaluewithoutcausing
the “FAC” icon to appear, your conductivity or pH/ORP sensor is 
damaged and must be replaced.
18










