NOTES: If a wrong buffer is added (outside of 6-8 pH),“7” and “BUFFER” 
willash,andtheDialysateMeterwillnotadjust.
The uncalibrated pH value displayed in step 4 will assist in determining 
the accuracy of the pH sensor.  If  the  pH  reading  is  above  8  with  pH 
7  buffer  solution,  the  sensor  well  needs  additional  rinsing  or  the  pH 
sensor is defective and needs to be replaced
.
5.  Press  or  until the display reads 7.00.
NOTE: Attempted calibration of >1 pH unit from factory calibration will 
cause “FAC” to appear. This indicates the need for sensor replacement 
(ref. Troubleshooting pg. 37-38) or fresh buffer solution. 
WARNING: The “FAC” internal electronic calibration is not intended to 
replace calibration with pH buffers. It assumes an ideal pH sensor. 
You  may  press  to accept the preset factory value, or 
you may
reduce your variation from factory setting by pressing  or 
.
6.  Press  to accept the new value. The pH Zero Calibration 
  is now complete. Continue with pH Gain Calibration. DO NOT  
  press any measurement keys as this will exit calibration.
      b. pH Gain Calibration (D-6)
Important: Always calibrate or verify your Dialysate Meter with a pH 7 
buffer solution before adjusting the gain with acid or base buffers, i.e., 
4 and/or 10, etc. Either acid or base solution can be used for the 2nd 
point  “Gain”  calibration  and  then  the  opposite  for  the  3rd  point.  The 
display will verify that a buffer is in the sensor well by displaying either 
“Acd” or “bAS”.
1.  The pH calibration mode is initiated by either completion of 
  the pH Zero Calibration, or verifying 7 buffer and pressing the 
  key twice while in pH measurement mode.
2.  At this point the “CAL”, “BUFFER” and “Acd” or “bAS”
  annunciators will be displayed (see Figures 6 and 7).
NOTE: If the “Acd” and “bAS” indicators are blinking, the unit is indicating an 
error and needs either an acid or base solution present in the sensor well
.