Manual
  B. Calibration Tracking Records for General Use (Not Dialysate) 
To  minimize  your  calibration  effort,  keep  records.  If  adjustments  you 
are making are minimal for your application, you can check less often. 
Changes in conductivity  calibration should  be recorded  in percent. 
Changes in pH calibration (D-6) are best recorded in pH units.
Calibration is purposely limited in the Dialysate Meter to 
±
10% for the 
conductivity cell, as any change beyond that indicates damage, not 
drift. Likewise, calibration changes are limited to 
±
1 pH unit, as any 
change beyond that indicates the end of the sensor’s lifetime and 
replacement is recommended.
WARNING:  For Dialysate measurements, it is critical that you verify that 
the pH sensor is working at the beginning and end of each workday 
or more often if required by your clinic’s internal procedures. 
  C. Conductivity, RES, TDS Practices to Maintain Calibration
Note:  Very active chemicals may discolor the conductivity electrodes 
(though this is not likely in clinical use). 
• Thisdoesnotaffecttheaccuracyofconductivitymeasurements.
1. Cleanoilylmsororganicmaterialfromthecellelectrodesby 
  pouring isopropyl alcohol or squirting a foaming non-abrasive 
  cleaner into the cell and rinsing (see Cleaning Sensors, pg. 34).
• DO NOT scrub inside the cell. 
2.  Calibrate with solutions close to the measurements you make.  
  Readings are compensated for temperature based on solution  
  type. If you choose a solution type that does not closely match  
  the characteristics of the solution you are trying to measure, your 
 recordsofcalibrationchangeswillreecttemperaturechanges 
  more than changes in the instrument’s accuracy. For example, 
  if you choose to measure tap water solute concentration in KCl  
  mode, but you calibrate with 442 because it is handy, the further  
  away from 25ºC the tap water sample is, the more error there  
  will be in the temperature-compensated reading. 
Note: Dialysate  
  conductivity measurements MUST be set to NaCl.
3.  Rinse out the cell with pure water after taking measurements.
• Allowingslowdissolvingcrystalstoforminthecell
contaminates future samples. 
4.  For maximum accuracy, keep the pH sensor cap on tight so that  
 nouidinthepHsensorwellwashesintotheconductivitycell.
  D. pH and ORP Practices to Maintain Calibration (D-6)
1.  ALWAYS keep the sensor wet with Myron L pH Sensor 
  Storage Solution when not in use. If the pH/ORP sensor is 
 allowedtodryout,itsaccuracywillbesignicantly 
  compromised.
24










