Manual
2. RinseandllpH/ORPsensorwellwithpHSensorStorage 
  Solution immediately after each use.
• NEVER use distilled water or tap water as a substitute for pH 
Sensor Storage Solution. This will deplete the sensor quickly, 
signicantlydecreasingitslifespan.
• IfpHSensorStorageSolutionisunavailable,use:
• AnalmostsaturatedKClsolution(preferred)
• AsaturatedsolutionofNaCl(tablesalt)andtapwater
(next best alternative)
• pH4buffer(leastdesirablealternative).
• ReplacewithMyronLpHSensorStorageSolutionassoon
as possible.
• MakesurethepH/ORPsensorcapisreinstalledsecurely
over the pH/ORP sensor well.
ORP  calibration  solutions  are  caustic,  and 
±
5% is  considered  very 
accurate.  By  using  the  pH  zero  setting  (0  mV  =  7  pH)  for  ORP  and 
precision electronics for detection, the D-6 delivers better accuracy without 
calibration than a simpler instrument could using calibration solutions. 
XII.  MEMORY
This feature allows up to 100 readings with their temperatures to be stored 
simultaneously for later recall. At the same time, the TIME and DATE are 
also recorded. Memory download to a computer requires the bluDock™ 
accessory package. Order MODEL # BLUDOCK. 
  A. Memory Storage
1.  While displaying a
  measurement, press   
  to record the   
  displayed value.  
2.  “MEMORY” will appear 
  and the temperature 
  display will be momentarily replaced by a number (1-100) 
  showing the position of the record. Figure 8 shows a reading 
 of1806µSstoredinmemoryrecord#4.
  B. Memory Recall 
1.  Press any measurement key.
2.  Press  , “MEMORY” will appear, and the display will show
  the last record stored.
Figure 8
°C
KCl
COND
MEMORY
25