Manual
2. RinseandllpH/ORPsensorwellwithpHSensorStorage
Solution immediately after each use.
• NEVER use distilled water or tap water as a substitute for pH
Sensor Storage Solution. This will deplete the sensor quickly,
signicantlydecreasingitslifespan.
• IfpHSensorStorageSolutionisunavailable,use:
• AnalmostsaturatedKClsolution(preferred)
• AsaturatedsolutionofNaCl(tablesalt)andtapwater
(next best alternative)
• pH4buffer(leastdesirablealternative).
• ReplacewithMyronLpHSensorStorageSolutionassoon
as possible.
• MakesurethepH/ORPsensorcapisreinstalledsecurely
over the pH/ORP sensor well.
ORP calibration solutions are caustic, and
±
5% is considered very
accurate. By using the pH zero setting (0 mV = 7 pH) for ORP and
precision electronics for detection, the D-6 delivers better accuracy without
calibration than a simpler instrument could using calibration solutions.
XII. MEMORY
This feature allows up to 100 readings with their temperatures to be stored
simultaneously for later recall. At the same time, the TIME and DATE are
also recorded. Memory download to a computer requires the bluDock™
accessory package. Order MODEL # BLUDOCK.
A. Memory Storage
1. While displaying a
measurement, press
to record the
displayed value.
2. “MEMORY” will appear
and the temperature
display will be momentarily replaced by a number (1-100)
showing the position of the record. Figure 8 shows a reading
of1806µSstoredinmemoryrecord#4.
B. Memory Recall
1. Press any measurement key.
2. Press , “MEMORY” will appear, and the display will show
the last record stored.
Figure 8
°C
KCl
COND
MEMORY
25