Manual

• Ifthe pHreading isbetween8.0pHand6.0pH,calibratingthe
sensor (see pH Calibration, pg. 20) will correct the offset.
7. Press to return the instrument to measurement mode
without changing the calibration of the instrument.
3. pH/ORP
CAUTION: ONLY authorized members of your facility’s maintenance
staff should attempt the following procedures.
TheuniquepH/ORPsensorinyourDialysateMeterisanonrellable
combination type that features a porous liquid junction.
It should not be
allowed to dry out.
However, if this occurs, the sensor may sometimes
berejuvenatedbyrstcleaningthesensorwellwithisopropylalcohol
or a liquid spray cleaner, such as Windex™ or Fantastic™, and rinsing
well. Do not scrub or wipe the pH/ORP sensor.
Then use one of the following methods:
1. Pour a HOT salt solution ~60°C/140°F — preferably potassium
chloride (KCI) solution (Myron L pH/ORP Sensor Storage
Solution) but HOT tap water with table salt (NaCl) will work
ne—inthepH/ORPsensorwellandallowtocool.Retest.
or
2. Pour DI water in the pH/ORP sensor well and allow to stand for
no more than 4 hours (longer can deplete the reference
solution and damage the glass bulb). Retest.
If neither method is successful, the sensor must be replaced.
DriftingcanbecausedbyalmonthepHsensorbulband/or
reference (see Figure 31). Use isopropyl alcohol or spray a liquid
cleaner such as Windex™ or Fantastic™ into the sensor well to clean
it. The sensor bulb is very thin and delicate. Do not scrub or wipe the
pH/ORP sensor.
pH/ORP SENSOR
Top View
ORP
Electrode
pH Glass
Electrode
Sensor
Body
Reference
Junction
under Glass
pH Bulb
Figure 31
35