Manual
XX. TROUBLESHOOTING CHART
37
SYMPTOM POSSIBLE CAUSE
Meter won’t turn ON •Batteryweak.
Inaccurate pH readings
(D-6)
•pHcalibrationneeded.
•ContaminationfromresidualpHbuffers
or samples in sensor well.
•CalibrationwithexpiredpHbuffers.
No response to pH
changes (D-6)
•pH/ORPsensorbulbiscracked.
•Electromechanicalshortinsidethe
sensor body.
pH Zero Point calibration
will not adjust down to pH
7.0 (D-6)
•pH/ORPsensor’sinternalKClsolution
is depleted or contaminated.
pH readings drift or
respond slowly to changes
in buffers/samples (D-6)
or
“FAC” is displayed
repeatedly
•TemporaryconditionduetothepH/
ORP sensor’s memory of solution in left
in the sensor well for long periods.
•Bulbdirtyordriedout.
•Referencejunctioncloggedorcoated.
Unstable Conductivity or
TDS reading
•Dirtyelectrodes.
Unstable Resistivity
readings
•Dirtyelectrodes.
•Contaminationofsampledueto
exposure to air resulting in samples
withresistivity>1MΩ.
Unable to calibrate
Conductivity or TDS
•Filmordepositsonelectrodes.
Resistivity readings much
lower than expected
•Contaminationfromprevioussampleor
from pH sensor well.
•TestSampleiscontaminatedby
Carbon dioxide absorbed from the air
(commonly caused by using a standing
sample).