Manual
38
CORRECTIVE ACTION
•Checkbatteryconnectionsorreplacebattery.
(See Battery Replacement, pg. 32.)
•RecalibratepHfunction.(SeepHCalibration(D-6),pg.20.)
•Thoroughlyrinsesensorwell.
•Recalibrateusingfreshbuffers.
(See pH Calibration (D-6), pg. 20.)
•ReplacepH/ORPSensor.
(See pH/ORP Sensor Replacement (D-6), pg. 33.)
•Cleanandrejuvenatesensor.(SeepH/ORP,pg.35.)
•Recalibrate.(SeepHCalibration(D-6),pg.20.)
•Ifnoimprovement,replacepH/ORPsensor.
(See pH/ORP Sensor Replacement (D-6), pg. 33.)
•Cleanandrejuvenatesensor.(SeepH/ORP,pg.35.)
•Recalibrate.(SeepHCalibration(D-6),pg.20.)
•Ifnoimprovement,replacepH/ORPsensor.
(See pH/ORP Sensor Replacement (D-6), pg. 33.)
•Cleancellcupandelectrodes.(SeeCleaningSensors,pg.34.)
•Minimizetestsampleexposuretoairbyusingaowingsample.
Letsampleowcontinuouslyintoconductivitycellwithnoaeration.
•Cleancellcupandelectrodes.(SeeCleaningSensors,pg.34.)
•Minimizetestsampleexposuretoairbytakingaowingsample.
Letsampleowcontinuouslyintoconductivitycellwithnoaeration.
•Cleancellcupandelectrodes.(SeeCleaningSensors,pg.34.)
•Rinsecellcupmorethoroughlybeforemeasurement.
•EnsurepHcapissnuglyinplace.
•Minimizetestsampleexposuretoairbytakingaowingsample.
Letsampleowcontinuouslyintoconductivitycellwithnoaeration.