WARNING: The pH/ORP sensor well must ALWAYS be kept wet with
a pH sensor storage solution when not in use.
• Allowingthesensortodryoutwilldamageorruinit.
• NEVER use distilled water or tap water as a substitute for pH
sensor storage solution.
WARNING: DO NOT use a Digital Dialysate Meter™ in place of the
primary sensor controls and/or monitors of the hemodialysis system.
• IfmeasurementstakenwithaDigitalDialysateMeter™differ
signicantlyfromthosereportedbythehemodialysissystem:
• DO NOT continue with dialysis treatment.
• IMMEDIATELY investigate the cause of the variation.
WARNING: Perform calibrations and calibration checks with standard
reference solutions.
• MyronLCompanyreferencesolutionsarehighlyrecommended
and are the same solutions used in the manufacture and factory
calibration of your Digital Dialysate Meter™.
WARNING: When calibrating the conductivity/resistivity or TDS
functions of a Digital Dialysate Meter™, press and release the CAL
key.
• DO NOT press and hold the CAL key. This may inadvertently
change the conductivity/resistivity/TDS solution type and alter the
performance of the instrument.
CAUTION: NEVER store or operate a Digital Dialysate Meter™
outside of its operating/storage temperature limits (see General
Specications,pg.6).
• DO NOT put solutions below 0°C/32°F or in excess of 71°C/160°F
in the sensor area.
CAUTION:If“LOBATT”ashes,replacethe9Valkalinebattery.Follow
the instructions in the Digital Dialysate Meter™ Operation Manual
precisely.
CAUTION: Federal law restricts this device to sale by or on the order
of a physician.