1.  Turn the Range Switch to the desired range. If you don’t 
  know which range to use, set it to 1000.
2.  Rinse the cell cup three times with the sample you want 
  to test. (For very hot or very cold samples see 
  TEMPERATURE COMPENSATION below.)
NEVER FILL THE CELL BY DIPPING THE METER INTO 
WATER!
3.  Fill the cell with another sample to at least 1/4” (6mm) 
  above the upper electrode.
4.  Press the black button.
5.  Read the dial’s black scale value indicated by the 
  pointer. If the pointer is very low on the left or goes off 
  the scale to the right, try the next lower or higher range.
6. Multiply the dial’s black scale value by the range setting 
  to determine conductivity in micromhos (microsiemens).
7.  If the pointer goes off the scale to the right on the x1000 
  range, use an RE-10 range extender (see 
  ACCESSORIES).
NOTE:
When you are nished with the meter, RINSE THE CELL 
CUP with clean water, preferably distilled or deionized.
TEMPERATURE COMPENSATION: For very hot or very 
cold solutions, let the three rinse samples each remain in 
the cell for several seconds. Then immediately ll the cell  
with the sample you want to test (step 3 above). This 
allows the automatic temperature compensation feature 
time to work properly.
RESISTIVITY MEASUREMENT: Model EP only. Accurate 
measurement of high purity water is difcult even under  
controlled conditions. Water temperature and even carbon 
dioxide (CO
2
) in the atmosphere will greatly inuence  
resistivity. Use the following 
methods for best accuracy:
1.  Turn the range switch to the x.1 range (with the “Ω” 
  symbol around it).
2.  Hold your meter close to the high purity water outlet. 
  While pressing the button, allow the water to ow into 
  the Cell Cup and out again, continuously purging the 
  cell. Do not allow air bubbles to collect.
3.  As the cell is washed clean of trace contamination, the 
  reading will drop and eventually stabilize. Read the 
  resistivity directly in megohms on the red scale. DO 
  NOT multiply the dial reading by the range switch value.
NOTE:
The closer the tested water temperature is to 25ºC 
(77ºF), and the less it is exposed to air, the more accurate 
your reading will be.
CAUTION
ALWAYS rinse the cell cup with pure water immediately 
after use or cleaning. This will prevent the build-up of 
unwanted deposits and possible contamination of the next 
sample.
DO NOT use with samples hotter than specied. The  
readings WILL NOT be accurate.
DO NOT splash solvents such as lacquer thinner, acetone, 
benzene or chlorinated solvents on the plastic case.
DO NOT x or modify the meter. That will void your  
warranty. See SERVICE for details or consult Myron L 
Company
DO NOT DIP THE INSTRUMENT INTO WATER. If water 
does get inside the instrument, see MAINTENANCE for 
instructions on drying it.
MODEL EP
MODEL EP
MODEL EP10
USING YOUR METER