User Manual
14
15
  4. FC
E
 Best Practices 
For best results it is recommended that you:
1.  Take  3  consecutive  FC
E
  measurements  and  record  the 
readings.
2.  Calculate the average of the 3 measurements. Use this value.
3.  Ignore  measurements  that  are  signicantly  different  from  the 
others. Ex: 3.20 ppm, 1.15 ppm, 3.10 ppm
IMPORTANT:    After  use,  ll  pH/ORP  sensor  well  with  Myron  L  pH 
Sensor  Storage  Solution  and  replace  protective  cap.  If  Myron  L  pH 
Sensor  Storage  Solution  is  unavailable,  you  can  use  a  strong  KCl 
solution,  a  pH  4  buffer,  or  a  saturated  solution  of  table  salt  and  tap 
water  (ref.  Cleaning  Sensors,  2.  pH/ORP,  pg.  32).  Do  not  allow  pH/
ORP sensor to dry out. 
VI.  SOLUTION SELECTION
  A. Why Solution Selection is Available
Conductivity, MIN/SALT, and  TDS require  temperature correction  to 
25°C values (ref. Standardized to 25°C, pg. 37). Selection determines 
the temperature correction of conductivity and calculation of TDS from 
compensated conductivity (ref. Cond. Conversion to TDS, pg. 40).
  B. The 3 Solution Types
On the left side of the display is the salt solution characteristic used 
to  model  temperature  compensation  of  conductivity  and  its  TDS 
conversion.  Generally,  using  KCl  for  Conductivity,  NaCl  for  Mineral/
Salt, and 442 (Natural Water characteristic) for TDS will reect present 
industry practice for standardization. This is the setup as shipped from 
the factory (ref. Solution Characteristics, pg. 40).
  C. Calibration of Each Solution Type
There is a separate calibration for each of the 3 solution types. Note 
that  calibration  of a  442  solution  does  not  affect  the  calibration of a 
NaCl solution. For example: Calibration (ref. Conductivity, MIN/SALT 
or  TDS  Calibration,  pg.  17)  is  performed  separately  for  each  type 
of  solution  one  wishes  to  measure  (ref.  Conductivity/TDS  Standard 
Solutions, pg. 36).










