User Manual
16
17
The
 becomes an “ACCEPT” key. At each point, pressing 
accepts the new calibration value and steps you to the next adjustment 
(or out of CAL mode if there are no more steps). 
To bypass a calibration step, just press   to accept the present 
value as is.
      b. Leaving Calibration
Calibration  is  complete  when  the  “CAL”  icon  goes  out.  Pressing  any 
measurement key cancels changes not yet accepted and exits calibration 
mode. 
Leaving pH after the 2nd buffer results in the same gain being entered 
in place of the 3rd buffer.
    2. Calibration Limits
There  are  calibration  limits. A  nominal  “FAC”  value  is  an  ideal  value 
stored  by  the  factory. Attempts  to  calibrate  too  far,  up  or  down,  from 
there will  cause the displayed value  to  be replaced with “FAC”. If you 
accept it (press the “Cal” key), you will have the original default factory 
calibration for this measurement. The need to calibrate so far out that 
“FAC” appears indicates a procedural problem, wrong standard solution, 
a very dirty cell cup or an aging pH/ORP sensor (ref. Troubleshooting 
Chart, pg. 34).
  C. Calibration Procedures 
    1. Conductivity, MIN/SALT or TDS Calibration
a.  Rinse conductivity cell three times with proper standard (KCl, 
  NaCl, or 442) (ref. Cond/TDS Standard Solutions, pg. 36). 
b.  Rell conductivity cell with same standard. NACL-7500 shown. 
c.  Press
, 
MIN
SALT
 or
, then 
  press
, “CAL” icon will 
  appear on the display 
  (see Figure 4). 
d.  Press or to
 or
  to step the displayed value toward the standard’s value (7582
Figure 4 
°C
NaCl
COND
CAL
µS
7582
23.8










