User Manual
34
35
XVII.  TROUBLESHOOTING CHART
Symptom Possible Cause Corrective Action
No display, even though 
measurement key pressed
Battery weak or not connected. Check connections or replace battery. Ref. Battery Replacement, pg. 31.
Inaccurate pH readings 1.  pH calibration needed. Ref. pH Cal., 
 pg. 18.
2.  Cross-contamination from residual pH 
 buffers or samples in sensor well.
3.  Calibration with expired pH buffers.
1.  Recalibrate instrument.
2.  Thoroughly rinse sensor well.
3.  Recalibrate using fresh buffers. Ref. pH Buffer Solutions, pg. 36.
No response to pH changes Sensor bulb is cracked or an 
electromechanical short caused by an 
internal crack.
Replace pH/ORP sensor. Ref. Replacement pH/ORP Sensor, pg. 37.
Will not adjust down to pH 7 pH/ORP sensor has lost KCl. Clean and rejuvenate sensor (ref. Cleaning Sensors, pg. 32) and recalibrate. If no 
improvement, replace pH/ORP sensor (ref. Replacement pH/ORP Sensor, pg. 37). 
pH readings drift or respond slowly to 
changes in buffers/samples or “FAC” 
is displayed repeatedly
1.  Temporary condition due to “memory” of 
 solution in pH sensor well for long periods.
2.  Bulb dirty or dried out.
3.  Reference junction clogged or coated.
Clean and rejuvenate sensor (ref. Cleaning Sensors, pg. 32) and recalibrate. If no 
improvement, replace pH/ORP sensor (ref. Replacement pH/ORP Sensor, pg. 37).
Unstable Conductivity/TDS/ 
MIN/SALT readings
Dirty electrodes. Clean cell cup and electrodes. Ref. Cleaning Sensors, pg. 32.
Unable to calibrate Conductivity/
TDS/MIN/SALT
Film or deposits on electrodes. Clean cell cup and electrodes. Ref. Cleaning Sensors, pg. 32.
Low ORP Reading
Slow or no response to ORP changes
ORP platinum electrode is dirty. Check the ORP sensor functioning. Take an ORP reading of Myron L pH/ORP 
Sensor Storage Solution (ref. pH Sensor Storage Solution, pg. 36. If the reading is 
outside the range of 350-400 mV, clean ONLY the platinum ORP 
electrode with Myron L ORP Conditioner solution-soaked cotton swab (ref. ORP 
Sensor Conditioner Solution, pg. 36), being careful not to touch the swab to the 
glass bulb of the pH sensor.
FC
E 
responds very slowly or returns an 
atypically high Predictive ORP value
1.  Dirty platinum electrode (see above). 
2.  ORP sensor memory/battery effect. 
 Some ORP sensors exhibit a residual 
 charge when measuring LOW Free 
 Chlorine concentrations soon after 
 measuring a HIGH Free Chlorine 
 concentration.
1.  Rinse the pH/ORP sensor well briey with a small amount of ORP Sensor  
 Conditioner Solution. DO NOT leave the conditioning solution in the sensor well 
 for more than 10 seconds.
2.  Rinse the pH/ORP sensor 3 times with Sensor Storage Solution.
3.  Fill the sensor well with Sensor Storage Solution and let rest for 5 minutes.










