User Manual
44
45
  c. Temperature Compensation
pH sensor glass changes its sensitivity slightly with temperature, so the 
further  from  pH  7  one  is,  the  more  effect will  be  seen. A pH  of  11 at 
40°C would  be off by  0.2  units. The  Po o l Pr o   senses  the  sensor  well 
temperature and compensates the reading. 
  B. ORP/Oxidation-Reduction Potential/REDOX 
   1. ORP as an Indicator 
ORP is  the  measurement  of  the  ratio  of  oxidizing  activity  to  reducing 
activity in a solution. It is the potential of a solution to give up electrons 
(oxidize other things) or gain electrons (reduce). 
Like acidity and alkalinity, the increase of one is at the expense of the 
other, so  a  single  voltage  is  called  the  Oxidation-Reduction  Potential, 
with  a  positive  voltage  showing,  a  solution  wants  to  steal  electrons 
(oxidizing  agent).  For  instance,  chlorinated  water  will  show  a  positive 
ORP value.
   2. ORP Units
ORP is measured in millivolts, with no correction for solution temperature.  
Like pH, it is not a measurement of concentration directly, but of activity 
level.  In  a  solution  of  only  one  active  component,  ORP  indicates 
concentration. Also, as with pH, a very dilute solution will take time to 
accumulate a readable charge. 
   3. The ORP Sensor
An  ORP  sensor  uses  a  small  platinum  surface  to  accumulate  charge 
without  reacting  chemically.  That  charge  is  measured  relative  to  the 
solution, so  the solution  “ground”  voltage comes  from a  reference 
junction - same as the pH sensor uses.
   4. The Myron L ORP Sensor
Figure  31,  pg.  43,  shows  the  platinum  button  in  a  glass  sleeve.  The 
same  reference  is  used  for  both  the  pH  and  the  ORP  sensors.  Both 
pH  and  ORP will  indicate  0  for  a  neutral  solution.  Calibration  at  zero 
compensates for error in the reference junction. 
A  zero  calibration  solution  for  ORP  is  not  practical,  so  the  Po o l Pr o  
uses the offset value determined during calibration to 7 in pH calibration 
(pH 7 = 0 mV). Sensitivity of the ORP surface is xed, so there is no 
gain adjustment either.
   5. Sources of Error
The  basics  are  presented  in  pH/ORP,  pg.  42,  because  sources  of 
error are much the same as for pH. The junction side is the same, and 
though the platinum surface will not break like the glass pH surface, 
its  protective  glass  sleeve  can  be  broken.  A  surface  lm  will  slow 
the response time and diminish sensitivity. It can be cleaned off with 
detergent or acid, as with the pH glass. 










