Instruction Manual

1
1. Rinsethepen.Ifmeasuringfromacontainer,submergethepenandswirlitaround
inFRESHsamplesolution3times.Alternatively,swirlfor30secondsunder
astreamorinabodyofwater.
2. IftheMeasurementscreenisnotdisplayed,taptheMeasurement
buttonintheFeatureNavigationBar.Thescreenwillsay,
“Paired ULTRAPEN is Ofine“.
3. Graspthepennearthepencaptoavoidsamplecontamination.
4. Removepenfromtherinsesolution.
5. PressandreleasethePTBT2’spushbuttontoturnitON.
6. WhentheLEDashesrapidly,IMMEDIATELYdipthepeninthesamplesolutionso
thatthecelliscompletelysubmerged.Ifyoudonotsubmergethecellinsolution
beforetheashingslows,allowthepentopoweroffandretakethereading.
7. TheAppscreenwilldisplay“Measuring ...
8. WhentheLEDashesslowly,swirlthepentoremovebubbles,keepingthecell
submergedandpreferablyatleast1inch(2½cm)awayfromsides/bottomof
container.
a.InHoldmodewhentheLEDturnsonsolidthedisplaywillshowthefinalpHand
temperaturereadings.
b.InLIVEmodeallowthepentoremaininsolutionwhiletheLEDashesslowly.
ThedisplaywillshowlivepHandtemperaturereadings.Thepenwilltimeout
after5minute.Pushandreleasethepen’sbuttontopoweritOFFatany
timeduringLIVEmeasurement
9. TheULTRAPENwillturnOFFoncethemeasurementiscompleted.
V. Recording and Recalling Measurements
1. Onceameasurementisfinished,taptheMS button(MemoryStore).
2. TheAppscreenwilldisplaythemeasurementasanunsavedrecord.
3. TaptheSavebuttonintheupperrightcorneroftherecorddisplay.
4. TherecordwillbesavedandtheAppwillreturntothe
Measurementscreen.
5. TaptheMR button(MemoryRecall)intheFeatureNavigationBarto
displayalistofstoredrecords.
6. Swipeupanddowntorevealthefulllist.Tapanyrecordonthelist
toviewit’sdetails.
VI. Calibration
The Myron L
®
Company recommends calibrating twice a monthdepending on usage.
However, you should check the calibration whenever measurements are not as
expected.3-pointWetCalibrationismostaccurateandisrecommended.
NOTE:IfthemeasurementisNOTwithincalibrationlimits,“Error. Clean sensor. Check
Solution.”willdisplay.ChecktomakesureyouareusingaproperpHbuffersolution.If
thesolutioniscorrect,cleantheglassbulbofthesensorwithacottonswabsoakedin
isopropylalcohol.Restartcalibration.
NOTE:Smallbubblestrappedinthesensormaygiveafalsecalibration.
NOTE: If at any point during calibration, you do not submerge the sensor in solution
beforetheashingslows,allowthepentopoweroffandstartover.
A. Calibration Preparation
1. Formaximumaccuracy,fill2cleancontainerswitheachpHbufferyouwillbeusing.
Arrangetheminsuchawaythatyoucanclearlyrememberwhichistherinsesolution
and which is the calibration buffer. Always rinse the pH sensor between buffer
solutions.
2. EnsurethepHsensoriscleanandfreeofdebris.
NOTE:Ifyoudonothaveenoughbuffer,youcanuse1containerofeachbufferfor
calibrationand1containerofcleanwaterforallrinsing.
B. Multi-Point Calibration (2 and 3 point Calibration).
Toperformamulti-PointCalibrationyoumustalwaysstartwith7.0pHbuffer.
1. TaptheCALbuttonintheFeatureNavigationBar.
TheAppscreenwillappearandsay,“Paired ULTRAPEN is Ofine”.
2. Graspthepennearthebatterytoavoidsamplecontamination.
3. Removepenfromtherinsesolution.
4. Pressandreleasethepen’spushbutton.TheCalibrationscreenwilldisplay,“Press
and hold the button on the ULTRAPEN to enter CAL mode.”
5. PressandHoldthepen’sbutton.TheLEDwillgotosolidONandtheApp’sCalibration
screenwillappear
6. TapthegreenCALbuttonontheCalibrationscreen.
7. Whenthepen’sLEDashesrapidly,IMMEDIATELYdip
thepenincalibration7.0pHbuffersothatthecellis
completelysubmerged.TheCalibrationscreenwilldisplay.“Calibrating pH ...”
3
Continued on page 4 ...