Instruction Manual

4 5
PTBT2TFM 08-14
SPECIFICATIONS
MYRON L
®
COMPANY
2450ImpalaDrive•Carlsbad,Ca92010-7226•Phone:+1-760-438-2021
E-Mail:CustomerService–info@myronl.com•TechnicalSupport–techquestions@myronl.com
Website:www.myronl.com
MADEINUSA ©2014 MYRON L
®
COMPANY
Do not over tighten.
AppRequirements: AppleiOS6oriOS7.OptimizedforiOS7.
MeasurementRange: 0.00–14.00pH;Temperature:0°-71°C/32-160°F
Accuracy(AfterWetCalibration): ±0.01pH;Temperature±0.1ºC/±0.1ºF
Repeatability: ±0.01pH;Temperature±0.1ºC/±0.1ºF
Resolution: 0.01pH;Temperature±0.1ºC/±0.1ºF
TimetoReadingStabilization: 10-30seconds
ActiveModePowerConsumption: ActiveMode137mA,SleepMode2µA
TemperatureCompensation: Automaticto25ºC
PhysicalDimensions: 17.15cmLx1.59cmDor6.75in.Lx.625in.D
Weight: 54gor1.94oz
CaseMaterial: AnodizedAircraftAluminumwithProtectiveCoating
BatteryType: Ntype,Alkaline1.5V
CalibrationStandardSolution: 4.0pH,7.0pH,10.0pH
Operating/StorageTemperature: 0-55ºC/32-131ºF
WaterResistance: IP67andNEMA6
EN61236-1:2006-AnnexA:
2008;1999/5/ECAnnexIII
Electrostaticdischargetocaseofinstrumentmay
causePTBT2tospontaneouslypoweron.Inthiscase,
thePTBT2willpoweroffafterseveralseconds
FCCID:T7VPAN17 TheBluetoothtransceiverdevicemeetstherequirementsfor
modulartransmitterapprovalasdetailedinFCCpublicNotice
DA00-1407.
Canada(IC),license:
IC:216Q-PAN17
TheBluetoothtransceiverdevicemeetstherequirementsfor
modulartransmitterapprovalasdetailedinRSS-GEN.
ORIGINAL
EDITED FOR USE
Minimum height 5 mm
QUICK REFERENCE INSTRUCTIONS
1. TurnonULTRAPENApp:TapULTRAPENicononmobiledevice’shomescreen.
2. TurnonPTBT2:Pressandreleasepushbuttononpen.
3. LEDashesrapidly:Dippeninsamplesolutionsocellistotallysubmerged.
4. LEDashesslowly:Swirlpentoremovebubbles,keepcellsubmerged.
5. LEDturnssolidONthenturnsOFF:Measurementiscompleteandreadingisdisplayed.
Feature Navigation Bar
ThisbuttonalwaysreturnstheApptotheMeasurementScreen.
Thisbuttondisplaysalistofsavedmeasurements.
ThisbuttonplacestheAppinCalibrationmode.
ThisbuttonopenstheApp’sBluetoothConnect/Disconnectscreen.
ThisbuttonopenstheApp’spreferencesandsettingsmode.
AninternetlinktofullOperationsManual(Helpbutton)isfoundhere.
ACCESSORIES
STANDARDSOLUTIONSFORCALIBRATION:TheULTRAPEN™PT2requirespH4,pH7,andpH10
buffersolutionsforwetcalibrationandpHSensorStorageSolutionforproperstorage.
— Order MODEL#s: PH4, PH7, PH10, and PHSS.
ReplacementpHSensor(withinstructions)— Order Model: RPT2
.
III. pH Sensor Replacement (Follow the instructions that come with your Replacement Sensor.)
OnlyRemove/ReplacethepHsensorinaCLEANandDRYenvironment!
ToremovethepHsensor:
1. Removethesoakercap;makesurethePTBT2(includingthepHsensor)isclean
anddry.Loosenthebatterytraytoallowpressureequalization.
2. ThenfirmlygraspthepHsensorbodyandslowlypullthepHsensorout.
ToinstallanewpHsensor:
1. LineupthealignmenttabsonthepHsensorwiththealignmentslotsonthePT2
unit.
2. GentlypushthepHsensorintoposition,closebatterycap.
CAUTION:
MYRON L
®
WARRANTY
Exceptingsensorandbattery,theMyronL
®
CompanyPTBT2pHULTRAPENiswar-
rantedtobefreefrommanufacturedefectfor1year.ThepHsencerhasasix(6)month
limittedwarrenty.