DAQ 6023E/6024E/6025E Multifunction I/O Devices User Manual
Table Of Contents
- 6023E/6024E/6025E User Manual
- Support
- Important Information
- Contents
- About This Manual
- Chapter 1 Introduction
- Chapter 2 Installation and Configuration
- Chapter 3 Hardware Overview
- Chapter 4 Signal Connections
- I/O Connector
- Analog Input Signal Overview
- Analog Input Signal Connections
- Analog Output Signal Connections
- Digital I/O Signal Connections
- Programmable Peripheral Interface (PPI)
- Power Connections
- Timing Connections
- Field Wiring Considerations
- Chapter 5 Calibration
- Appendix A Specifications
- Appendix B Custom Cabling and Optional Connectors
- Appendix C Common Questions
- Appendix D Technical Support Resources
- Glossary
- Index
- Figures
- Figure 1-1. The Relationship Between the Programming Environment, NI-DAQ, and Your Hardware
- Figure 3-1. PCI-6023E, PCI-6024E, PCI-6025E, and PXI-6025E Block Diagram
- Figure 3-2. DAQCard-6024E Block Diagram
- Figure 3-3. Dithering
- Figure 3-4. CONVERT* Signal Routing
- Figure 3-5. PCI RTSI Bus Signal Connection
- Figure 3-6. PXI RTSI Bus Signal Connection
- Figure 4-1. I/O Connector Pin Assignment for the 6023E/6024E
- Figure 4-2. I/O Connector Pin Assignment for the 6025E
- Figure 4-3. Programmable Gain Instrumentation Amplifier (PGIA)
- Figure 4-4. Summary of Analog Input Connections
- Figure 4-5. Differential Input Connections for Ground Referenced Signals
- Figure 4-6. Differential Input Connections for Nonreferenced Signals
- Figure 4-7. Single Ended Input Connections for Nonreferenced or Floating Signals
- Figure 4-8. Single Ended Input Connections for Ground Referenced Signals
- Figure 4-9. Analog Output Connections
- Figure 4-10. Digital I/O Connections
- Figure 4-11. Digital I/O Connections Block Diagram
- Figure 4-12. DIO Channel Configured for High DIO Power-up State with External Load
- Figure 4-13. Timing Specifications for Mode 1 Input Transfer
- Figure 4-14. Timing Specifications for Mode 1 Output Transfer
- Figure 4-15. Timing Specifications for Mode 2 Bidirectional Transfer
- Figure 4-16. Timing I/O Connections
- Figure 4-17. Typical Posttriggered Acquisition
- Figure 4-18. Typical Pretriggered Acquisition
- Figure 4-19. SCANCLK Signal Timing
- Figure 4-20. EXTSTROBE* Signal Timing
- Figure 4-21. TRIG1 Input Signal Timing
- Figure 4-22. TRIG1 Output Signal Timing
- Figure 4-23. TRIG2 Input Signal Timing
- Figure 4-24. TRIG2 Output Signal Timing
- Figure 4-25. STARTSCAN Input Signal Timing
- Figure 4-26. STARTSCAN Output Signal Timing
- Figure 4-27. CONVERT* Input Signal Timing
- Figure 4-28. CONVERT* Output Signal Timing
- Figure 4-29. SISOURCE Signal Timing
- Figure 4-30. WFTRIG Input Signal Timing
- Figure 4-31. WFTRIG Output Signal Timing
- Figure 4-32. UPDATE* Input Signal Timing
- Figure 4-33. UPDATE* Output Signal Timing
- Figure 4-34. UISOURCE Signal Timing
- Figure 4-35. GPCTR0_SOURCE Signal Timing
- Figure 4-36. GPCTR0_GATE Signal Timing in Edge Detection Mode
- Figure 4-37. GPCTR0_OUT Signal Timing
- Figure 4-38. GPCTR1_SOURCE Signal Timing
- Figure 4-39. GPCTR1_GATE Signal Timing in Edge Detection Mode
- Figure 4-40. GPCTR1_OUT Signal Timing
- Figure 4-41. GPCTR Timing Summary
- Figure B-1. 68 Pin E Series Connector Pin Assignments
- Figure B-2. 68 Pin Extended Digital Input Connector Pin Assignments
- Figure B-3. 50 Pin E Series Connector Pin Assignments
- Figure B-4. 50-Pin Extended Digital Input Connector Pin Assignments
- Tables
- Table 3-1. Available Input Configurations
- Table 3-2. Measurement Precision
- Table 3-3. Pins Used by PXI E Series Device
- Table 4-1. I/O Connector Details
- Table 4-2. I/O Connector Signal Descriptions
- Table 4-3. I/O Signal Summary
- Table 4-4. Port C Signal Assignments
- Table 4-5. Signal Names Used in Timing Diagrams
Chapter 4 Signal Connections
© National Instruments Corporation 4-7 6023E/6024E/6025E User Manual
Table 4-3 shows the I/O signal summary for the 6023E, 6024E, and 6025E.
Table 4-3. I/O Signal Summary
Signal Name
Signal
Type and
Direction
Impedance
Input/
Output
Protection
(Volts)
On/Off
Source
(mA at V)
Sink
(mA
at V)
Rise
Time
(ns)
Bias
ACH<0..15> AI 100 GΩ
in
parallel
with
100 pF
42/35 — — — ±200 pA
AISENSE AI 100 GΩ
in
parallel
with
100 pF
40/25 — — — ±200 pA
AIGND AO — — — — — —
DAC0OUT
(6024E and 6025E only)
AO 0.1 Ω Short-circuit
to ground
5at10 5at-10 10
V/µs
—
DAC1OUT
(6024E and 6025E only)
AO 0.1 Ω Short-circuit
to ground
5at10 5at-10 10
V/µs
—
AOGND AO — — — — — —
DGND DO — — — — — —
VCC DO 0.1 Ω Short-circuit
to ground
1A fused — — —
DIO<0..7> DIO — V
cc
+0.5 13 at (V
cc
-0.4) 24 at
0.4
1.1 50 kΩ pu
PA<0 .. 7>
(6025E only)
DIO — V
cc
+0.5 2.5 at 3.7min 2.5 at
0.4
5 100 kΩ
pu
PB<0..7>
(6025E only)
DIO — V
cc
+0.5 2.5 at 3.7min 2.5 at
0.4
5 100 kΩ
pu
PC<0..7>
(6025E only)
DIO — V
cc
+0.5 2.5 at 3.7min 2.5 at
0.4
5 100 kΩ
pu
SCANCLK DO — — 3.5 at (V
cc
-0.4) 5at0.4 1.5 50 kΩ pu
EXTSTROBE* DO — — 3.5 at (V
cc
-0.4) 5at0.4 1.5 50 kΩ pu
PFI0/TRIG1 DIO — V
cc
+0.5 3.5 at (V
cc
-0.4) 5at0.4 1.5 50 kΩ pu
PFI1/TRIG2 DIO — V
cc
+0.5 3.5 at (V
cc
-0.4) 5at0.4 1.5 50 kΩ pu
PFI2/CONVERT* DIO — V
cc
+0.5 3.5 at (V
cc
-0.4) 5at0.4 1.5 50 kΩ pu
PFI3/GPCTR1_SOURCE DIO — V
cc
+0.5 3.5 at (V
cc
-0.4) 5at0.4 1.5 50 kΩ pu