Datasheet
Table 7. DC Accuracy/Resolution (Continued)
Output Type +6 V +25 V -25 V
Readback accuracy
16,18
± (% of reading + offset)
Voltage 0.1% + 5 mV 0.1% + 20 mV 0.1% + 20 mV
Current 0.2% + 10 mA 0.15% + 4 mA 0.15% + 4 mA
Programming resolution Voltage 1.7 mV 6.5 mV 6.5 mV
Current 0.30 mA 0.15 mA 0.15 mA
Readback resolution Voltage 0.41 mV 1.7 mV 1.7 mV
Current 70 µA 35 µA 35 µA
Load regulation
19
± (% of reading + offset)
Voltage 0.01% + 25 mV 0.03% + 5 mV 0.03% + 5 mV
Overvoltage protection 30 V (25 V channels) and 10 V (6 V channel)
Reverse voltage protection Reverse clamp diode, protected by
self-resetting fuse
External Trigger (TRIG)
Direction control
Input or output, software-selectable
Logic level 5 V compatible TTL input,
3.3 V LVTTL output
Drive strength 4 mA
Input voltage 0 V to 5 V
Note The external trigger line is designed to withstand accidental overvoltage from
signals on the VB-8012 or similar devices. It is not recommended for use with
signals likely to exceed 0 V to 5 V in normal operation.
Connectivity
Wired USB Interface
USB specification USB 2.0 Hi-Speed
18
Programming and readback accuracy specified at no load.
19
Change in output voltage for any load within range.
NI VirtualBench VB-8012 Specifications | © National Instruments | 11