Datasheet

Digital I/O
Type Pluggable screw terminal, 3.5 mm
(14 position)
Screw terminal wiring 0.1 to 2.0 mm
2
(30 to 14 AWG)
Torque 0.25 N · m (2.2 lb · in.)
Digital multimeter 4 mm banana jacks
DC power supply
Type Pluggable screw terminal, 3.81 mm
(6 position)
Screw terminal wiring 0.1 to 2.0 mm
2
(30 to 14 AWG)
Torque 0.25 N · m (2.2 lb · in.)
Security cable slot 1, complies with Kensington security slot
dimensions
If you need to clean the device, wipe it with a dry towel.
Safety Voltages
Connect only voltages that are within these limits.
DMM Isolation Voltages
Hazardous Voltage This icon denotes a warning advising you to take precautions
to avoid electrical shock.
Channel-to-earth ground
Continuous 300 V, Measurement Category II
Withstand 3,000 V
RMS
, verified by a 5 s dielectric
withstand test
Measurement Category II is for measurements performed on circuits directly connected to the
electrical distribution system. This category refers to local-level electrical distribution, such as
that provided by a standard wall outlet, for example, 115 V for U.S. or 230 V for Europe.
Caution Do not connect the VirtualBench hardware to signals or use for
measurements within Measurement Categories III or IV.
14 | ni.com | NI VirtualBench VB-8012 Specifications