DC Power Supply Isolation Voltages
+25 V and -25 V-to-earth ground, continuous 60 VDC, Measurement Category I
Note Measurement Categories CAT I and CAT O are equivalent. These test and
measurement circuits are not intended for direct connection to the MAINS building
installations of Measurement Categories CAT II, CAT III, or CAT IV.
Environmental
Operating temperature
0 °C to 40 °C
Storage temperature -20 °C to 70 °C
Operating humidity 10% to 90% RH, noncondensing
DMM full accuracy at 10% to 80%
Storage humidity 5% to 95% RH, noncondensing
Cooling Forced air circulation (positive pressurization)
through a fan. Fan speed automatically adjusts
according to operating conditions. Intake and
exhaust locations are on rear of device. Ensure
that the intake and exhaust locations are not
obstructed.
Pollution Degree 2
Maximum altitude 2,000 m
Indoor use only.
Shock and Vibration
Operational shock
30 g peak, half-sine, 11 ms pulse (Tested
in accordance with IEC 60068-2-27. Test
profile developed in accordance with
MIL-PRF-28800F.)
Random vibration
Operating 5 Hz to 500 Hz, 0.3 g
rms
Nonoperating 5 Hz to 500 Hz, 2.4 g
rms
(Tested in accordance
with IEC 60068-2-64. Nonoperating test
profile exceeds the requirements of
MIL-PRF-28800F, Class 3.)