Datasheet

Waveform Measurements
Oscilloscope time
7
Period, frequency, positive duty cycle,
negative duty cycle, positive pulse width,
negative pulse width, rise time, fall time,
rise rate, fall rate
Oscilloscope voltage
7
High, low, amplitude, maximum, minimum,
peak-to-peak, overshoot, undershoot, RMS,
mean, cycle RMS, cycle mean
Logic analyzer time
7
Period, frequency, positive duty cycle,
negative duty cycle, positive pulse width,
negative pulse width
Waveform Math
Operations
8
A + B, A - B, A * B, A/B, FFT
Function Generator (FGEN)
Waveforms
Sine, square, ramp/triangle, DC, arbitrary
9
Update rate 125 MS/s
Resolution 14 bits
Number of channels 1
Output impedance 50 Ω
Switchable filter
10
36 MHz lowpass, 7-pole, elliptical
Sine
Maximum frequency 20 MHz
Total Harmonic Distortion (THD)
1 MHz -55 dBc
10 MHz -50 dBc
Spurious Free Dynamic Range (SFDR) -70 dB at 1 MHz (non-harmonic)
Phase noise (1 MHz) -115 dBc/Hz at 10 kHz offset
7
Waveform measurements are only available in the VirtualBench application.
8
Waveform math is only available in the VirtualBench application.
9
Arbitrary waveforms are only available with the NI VirtualBench driver.
10
Switchable filters are only available with the NI VirtualBench driver. The VirtualBench application
automatically enables the lowpass filter in sine mode.
4 | ni.com | NI VirtualBench VB-8012 Specifications