DC and AC current
DMM A current connector fuse Internal ceramic fuse, 10 A 250 V, time-delay,
5 × 20 mm, T 10A H 250V (Bussmann part
number S505H-10-R at
www.cooperindustries.com)
DMM mA current connector fuse Internal ceramic fuse, 1.25 A 250 V,
time-delay, 5 × 20 mm, T 1.25A H 250V
(Bussmann part number S505H-1.25-R at
www.cooperindustries.com)
Caution Fuses are located on bottom of device underneath door. Use Phillips #1
screwdriver for removal. Ensure all hazardous voltages are disconnected from the
device prior to removal of door.
Fuse When this fuse symbol is marked on a device, take proper precautions.
Maximum common-mode voltage 300 V DC or AC
rms
DC
Table 2. DC Voltage Accuracy
Range Input Impedance
1-Year Accuracy
13
± (% of Reading +
% of Range)
Temperature Coefficient
13
± (% of Reading +
% of Range)/°C
100 mV
*
>10 GΩ, 10 MΩ 0.015 + 0.005 0.001 + 0.0005
1 V >10 GΩ, 10 MΩ 0.015 + 0.005 0.001 + 0.0005
10 V >10 GΩ, 10 MΩ 0.015 + 0.005 0.001 + 0.0005
100 V 10 MΩ 0.035 + 0.005 0.005 + 0.0005
300 V 10 MΩ 0.035 + 0.005 0.005 + 0.0005
* Add 15 µV if not immediately following offset null.
13
Indicates warranted specifications valid at T
cal
±5 °C. Temperature coefficients are calculated
using the temperature change from last external calibration.