User's Manual
© National Instruments Corporation 13 NI 4070/4072 Calibration Procedure
45. Set the input resistance of the NI 4070/4072 to >10 GΩ by calling 
niDMM_SetAttributeViReal64 with the following parameters:
• Attribute_ID = 
NIDMM_ATTR_INPUT_RESISTANCE
• Attribute_Value = NIDMM_VAL_GREATER_THAN_10_GIGAOHM 
46. Call 
niDMM_Read. Verify that this measurement falls between the 
limits listed in Table 15.
47. Set the input resistance of the NI 4070/4072 to 10 MΩ by calling 
niDMM_SetAttributeViReal64 with the following parameters:
• Attribute_ID = 
NIDMM_ATTR_INPUT_RESISTANCE 
• Attribute_Value = NIDMM_VAL_10_MEGAOHM 
48. Call 
niDMM_Read. Verify that this measurement falls between the 
limits listed in Table 15.
49. Output –1 V on the calibrator.
50. Set the input resistance of the NI 4070/4072 to >10 GΩ by calling 
niDMM_SetAttributeViReal64 with the following parameters:
• Attribute_ID = 
NIDMM_ATTR_INPUT_RESISTANCE
• Attribute_Value = NIDMM_VAL_GREATER_THAN_10_GIGAOHM 
51. Call 
niDMM_Read. Verify that this measurement falls between the 
limits listed in Table 15.
52. Set the input resistance of the NI 4070/4072 to 10 MΩ by calling 
niDMM_SetAttributeViReal64 with the following parameters:
• Attribute_ID = 
NIDMM_ATTR_INPUT_RESISTANCE
• Attribute_Value = NIDMM_VAL_10_MEGAOHM 
53. Call 
niDMM_Read. Verify that this measurement falls between the 
limits listed in Table 15.
54. Output 10 V on the calibrator.
55. Call 
niDMM_ConfigureMeasurement with the following 
parameters:
• Function = 
NIDMM_VAL_DC_VOLTS
• Range = 10
• Resolution = 10e–6
56. Set the input resistance of the NI 4070/4072 to >10 GΩ by calling 
niDMM_SetAttributeViReal64 with the following parameters:
• Attribute_ID = 
NIDMM_ATTR_INPUT_RESISTANCE
• Attribute_Value = NIDMM_VAL_GREATER_THAN_10_GIGAOHM 
57. Call 
niDMM_Read. Verify that this measurement falls between the 
limits listed in Table 15.