User's Manual
NI 4070/4072 Calibration Procedure 54 ni.com
• If you skip this section and you want to verify the new calibration 
coefficients before saving them to the EEPROM, repeat the 
Verification Procedures section (except for Initial Setup).
• If you skip this section and you do not want to verify the new 
calibration coefficients, go to the Completing the Adjustment 
Procedures section.
To adjust the current modes of the NI 4070/4072, complete the following 
steps:
1. Reset the calibrator.
2. Fasten the connectors on one end of the Fluke 5440 cable to the 
NI 4070/4072 HI SENSE and LO banana plug connectors, and fasten 
the connectors on the other end of the cable to the HI and LO calibrator 
binding posts. Figure 3 shows the correct connections. Table 9 lists the 
cable connections.
3. Call 
niDMM_ConfigureMeasurement with the following 
parameters:
• Function = 
NIDMM_VAL_DC_CURRENT
• Range = 0.02
4. Call niDMM_Read to configure the NI 4070/4072 for a current mode 
before applying current.
5. Output 20 mA on the calibrator with the current output set to 
NORM.
6. Call 
niDMM_CalAdjustGain with the following parameters:
• Mode = 
NIDMM_VAL_DC_CURRENT
• Range = 0.02
• Input Resistance = NIDMM_VAL_RESISTANCE_NA
• Expected Value = 0.02
7. Output –20 mA on the calibrator with the current output set to NORM.
8. Call 
niDMM_CalAdjustGain with the following parameters:
• Mode = 
NIDMM_VAL_DC_CURRENT
• Range = 0.02
• Input Resistance = NIDMM_VAL_RESISTANCE_NA
• Expected Value = –0.02
9. Output 0 A on the calibrator.
10. Call 
niDMM_CalAdjustOffset with the following parameters:
• Mode = 
NIDMM_VAL_DC_CURRENT
• Range = 0.02
• Input Resistance = NIDMM_VAL_RESISTANCE_NA