User's Manual
© National Instruments Corporation 21 NI cDAQ-9172 User Guide and Specifications
triggers can come from three separate modules if desired. To find your 
module triggering options, refer to the documentation included with your 
C Series I/O modules. For more information about using digital modules 
for triggering, refer to the Digital I/O section.
AI Start Trigger Signal
Use the AI Start Trigger (ai/StartTrigger) signal to begin a measurement 
acquisition. A measurement acquisition consists of one or more samples. If 
you do not use triggers, begin a measurement with a software command. 
Once the acquisition begins, configure the acquisition to stop in one of the 
following ways:
• When a certain number of points is sampled (in finite mode) 
• After a hardware reference trigger (in finite mode) 
• With a software command (in continuous mode) 
An acquisition that uses a start trigger (but not a reference trigger) is 
sometimes referred to as a posttriggered acquisition. That is, samples are 
measured only after the trigger. 
When you are using an internal sample clock, you can specify a delay from 
the start trigger to the first sample.
Using a Digital Source
To use ai/StartTrigger with a digital source, specify a source and an edge. 
Use the following signals as the source:
•Any PFI terminal.
• Counter n Internal Output
The source also can be one of several other internal signals on your 
NI cDAQ-9172 chassis. Refer to the Device Routing in MAX topic in the 
NI-DAQmx Help or the LabVIEW Help in version 8.0 or later for more 
information. 
The NI-DAQmx Help is available after installation from Start»
All Programs»National Instruments»NI-DAQ»NI-DAQmx Help. 
To view the LabVIEW Help, in version 8.0 or later, select Help»
Search the LabVIEW Help in LabVIEW. Alternately, to download 
the LabVIEW Help, go to 
ni.com/manuals.
You also can specify whether the measurement acquisition begins on the 
rising edge or falling edge of ai/StartTrigger.