User's Manual
NI cDAQ-9172 User Guide and Specifications 70 ni.com
You can configure the filter setting for each input independently. On power 
up, the filters are disabled. Figure 41 shows an example of a low to high 
transition on an input with its filter set to 125 ns (N = 5).
Figure 41. Filter Example
Enabling filters introduces jitter on the input signal. For the 125 ns and 
6.425 µs filter settings, the jitter is up to 25 ns. On the 2.56 ms setting, the 
jitter is up to 10.025 µs.
Prescaling
Prescaling allows the counter to count a signal that is faster than the 
maximum timebase of the counter. The NI cDAQ-9172 chassis offers 
8X and 2X prescaling on each counter (prescaling can be disabled). Each 
prescaler consists of a small, simple counter that counts to eight (or two) 
and rolls over. This counter can run faster than the larger counters, which 
simply count the rollovers of this smaller counter. Thus, the prescaler acts 
as a frequency divider on the Source and puts out a frequency that is 
one-eighth (or one-half) of what it is accepting.
Figure 42. Prescaling
Prescaling is intended for frequency measurement where the measurement 
is made on a continuous, repetitive signal. The prescaling counter cannot 
be read; therefore, you cannot determine how many edges have occurred 
since the previous rollover. You can use prescaling for event counting 
provided it is acceptable to have an error of up to seven (or one). You can 
use prescaling when the counter Source is an external signal. Prescaling
is not available if the counter Source is one of the internal timebases 
(80MHzTimebase, 20MHzTimebase, or 100kHzTimebase).
1 2 3 1  4  1 2 3 4 5 
PFI Terminal
Filter Clock 
(40 MHz) 
Filtered Input 
Filtered input goes high 
when terminal is sampled 
high on five consecutive 
filter clocks. 
External Signal 
Counter Value 
Prescaler Rollover 
(Used as Source 
by Counter) 
0 1 