Datasheet

www.ncte.de
Revision L
Nov 2014
Torque Sensor
Series 3000 and Series 4000
Modifications reserved. All details describe our products in general form only. They are not to be
understood as a guarantee of quality or durability and do not constitute any liability whatsoever.
Page 4
4. Available Options
PWM-signal output
Description Unit Value
Carrier frequency Hz 980
Signal at torque = zero % 50
Measurement range % 10…90
Error indication % 95
Calibration parameter %/Nm 40 / Measurement range
Current output
Description Unit Value
Signal at torque = zero mA 12
Measurement range mA 4 … 20
Calibration parameter mA/Nm 8 / Measurement range
Frequency output
Description Unit Value
Basic frequency kHz 60
Measurement range kHz 20 … 100
Calibration parameter kHz/Nm 40 / Measurement range
4.1 Optional Signal Outputs
In addition to the analog output signal the Series 3000 und Series 4000 can also be delivered with another
optional output signal as listed below.
Only the analog voltage output is calibrated by default. All other output signals are adjusted
according to the analog voltage output.