User's Manual

Data Portion Contents
-------------------------------------------------------------------
DATA01 .. 04 : Lamp Hour Meter(Normal mode) (second)
DATA05 .. 08 : Reserved
DATA09 .. 12 : Lamp Use Warning Starting Time(Normal mode) (second)
DATA13 .. 16 : Lamp Use Prohibited Time(Normal mode) (second)
Response: At the time of a failure
A3H 8CH 01H xxH 02H DATA01 DATA02 CKS
(*1) (*2) (*4) (*3)
Supplement:
Example for acquiring remaining lamp time (in terms of Normal mode values)
: Lamp Hour Meter (Normal mode)
DATA01 DATA02 DATA03 DATA04
30H 2AH 00H 00H : 10800 seconds
Calculator Procedure
1) Set calculator to HEX
2) Punch in DATA04 DATA03 DATA02 DATA01 (00002A30)
3) Change from HEX to Decimal - value will change to 10800 seconds
: Starting time for lamp usage warning message (in terms of Normal mode values)
DATA09 DATA10 DATA11 DATA12
00H DDH 6DH 00H : 7200000 second
Calculator Procedure
1) Set calculator to HEX
2) Punch in DATA12 DATA11 DATA10 DATA09 (006DDD00)
3) Change from HEX to Decimal - value will change to 7200000 seconds
Lamp remaining time (in terms of Normal mode values)
= (7200000 - 10800) / 3600 = 1997 hour
037-2. LAMP INFORMATION REQUEST 2
**********************************************************************
Function:
This command acquires lamp remaining amount.
Command:
03H 94H 00H 00H 00H 97H
Response: At the time of a success
23H 94H 01H xxH 05H DATA01 .. DATA05 CKS
(*1) (*2) (*3)
Data Portion Contents
-------------------------------------------------------------------
DATA01 .. 04 Reserved
DATA05 lamp remaining amount (100% to -10%)
Response: At the time of a failure
A3H 94H 01H xxH 02H DATA01 DATA02 CKS
(*1) (*2) (*4) (*3)