User Manual

10000891 REV A
16
What You Hear
What It Means
One long beep
The Remote Control successfully paired with the Trial
Stimulator.
Three regular beeps
The Remote Control did not find the Trial Stimulator.
Make sure the Trial Stimulator Pairing Light is still
blinking and try again.
Refer to the Troubleshooting section of this manual.
Four regular beeps
The Remote Control found more than one Trial
Stimulator ready for pairing. Only one Trial Stimulator
should be ready for pairing at a time.
Refer to the Troubleshooting section of this manual.
4.2 Lead Placement
4.2.1 Percutaneous Lead Placement
NOTE: If sedation or Monitored Anesthesia Care is used, it is recommended that the patient remain
communicative during needle and lead placement to help mitigate any risk of neural injury.
1. Prep the skin and drape the patient in a usual sterile fashion.
2. Confirm pre-operative antibiotic prophylaxis if deemed appropriate.
3. Inject a local anesthetic at the desired needle insertion site.
CAUTION: Needles not provided by Nevro Corp. may damage Nevro leads.
Remote Control
ON / OFF Button
Program Selection Button