UPGRADE NE-1000X FAMILY OF SYRINGE PUMPS Owner's manual

NE-1000X 5 www.SyringePump.com
Phase Function Rate Time Direction
8
LINR 0 ml/hr 00:10 Infuse
Phase Function Rate Time Direction
9
LINR 1000 ml/hr 00:00 Infuse
Phase Function Rate Time Direction
10
LINR 1000 ml/hr 00:10 Infuse
Phase Function Rate Time Direction
11
LINR 0 ml/hr 00:00 Infuse
Phase Function
12
RET
The above program would be entered from RS-232 with the following commands:
PHN 1
FUN RAT
RAT 100 MH
VOL 1
DIR INF
PHN 2
FUN INC
RAT 200
TIM 00:10
DIR INF
PHN 3
FUN JMP 08
PHN 4
FUN RAT
RAT 500 MH
VOL 5.0
DIR INF
PHN 5
FUN DEC
RAT 100.0
TIM 00:10
DIR INF
PHN 6
FUN JMP 08
PHN 7
FUN JMP 01
PHN 8
FUN LIN
RAT 0.0
TIM 00:10
DIR INF
PHN 9
FUN LIN
RAT 1000 MH
TIM 00:00
DIR INF
PHN 10
FUN LIN
RAT 1000 MH
TIM 00:10
DIR INF
PHN 11
FUN LIN
RAT 0.0 MH
TIM 00:00
DIR INF
PHN 12
FUN RET
Return to Phase following calling
Program function
Linearly ramp the infusion rate
from 1000 ml/hr to 0 ml/hr over 10
minutes
Linearly ramp the infusion rate
from 0 ml/hr to 1000 ml/hr over 10
minutes
Linear Ramping Sub-Routine