User Manual

1. GENERAL
1.6 Operator's Manual WEE-1000
Installation
Warning - continued
Before defibrillation, all persons must keep clear of the bed and must
not touch the patient or any equipment connected to the patient.
Failure to follow this warning may cause serious electrical burn,
shock or other injury.
WARNING
Do not install the EEG System Program into a personal computer
which is not specified by Nihon Kohden and connect it to the access
point.
- If the personal computer does not satisfy the performance
specifications and safety standards which are required by Nihon
Kohden, the patient and operator may get electrical shock.
- Nihon Kohden does not warrant if hardware and/or software
becomes defective after installation.
Only use the provided power cords. If different power cord is used, it
may cause electrical shock.
For electrical safety, equipotential grounding is required. Consult a
qualified biomedical engineer.
Connect only the specified instruments to the connectors or socket
marked with , by following the specified procedure. Otherwise,
electrical leakage current may harm the patient and operator.
Do not install the access point near a microwave oven. The
microwaves from the microwave oven may interfere with the radio
wave communication between the telemetry unit and access point.
Connecting to a Local Area Network
When connecting the access point and electroencephalograph with a
local area network, connect the access point and
electroencephalograph so that the access point and
electroencephalograph are electrically separated from the local area
network according to the IEC 60601-1-1 “Medical electrical equipment -
Part 1-1: General requirements for safety - Collateral standard: Safety
requirements for medical electrical systems”. Failure to follow this
warning may cause electrical shock to the patient and operator.
Check that there is no damage on the surface of the network cable. If
it is damaged, it may cause electrical shock to the patient and
operator.