User Manual

1. GENERAL
Operator's Manual WEE-1000 1.7
Battery
CAUTION
When connecting the cables, make sure that all components of the
wireless input unit turned off.
Only install the specified software in the electroencephalograph.
Otherwise the electroencephalograph may malfunction.
Do not install the telemetry unit and access point in a place where is
blocked by metal or concrete, or do not install the access point with
its antenna bent. Decreased radio wave causes frequent signal loss
between the telemetry unit and access point.
Do not give impact to the antenna. This may damage the access point
or cause access point malfunction.
Use the provided ZR adapter when installing the access point.
Otherwise, the access point may fall off and cause injury.
After installing the telemetry unit and access point, check that the
communication between the telemetry unit and access point is
correctly performed without any interference.
WARNING
Keep the battery away from fire. Do not heat the battery. Otherwise,
the battery explodes.
Do not immerse the battery in water or seawater. The battery heats up
and rusts and the battery liquid leaks out.
Never use a battery which is damaged, discolored or has leakage. A
damaged battery explodes if used. If the battery is damaged and the
battery liquid contacts the eyes or skin, wash immediately and
thoroughly with water and see your physician. Never rub your eyes,
otherwise you may lose your eyesight.
Never disassemble, modify or give impact to the battery. The battery
short-circuits and the battery liquid leaks out.
Never short-circuit the + and – terminals on the battery with a wire.
Do not leave the battery where patients can reach it. If a battery is
swallowed, see your physician immediately.