2. INSTALLATION/PREPARATION
2.2 Operator's Manual WEE-1000
CAUTION
Select a room with a 3-prong outlet with a ground third contact.
Do not install the system near equipment with a high power
consumption, such as large X-ray equipment.
Do not install the system near a power line, dynamo or motor which
has electromagnetic induction.
Do not install the wireless input unit near an electrosurgical unit or
RF therapeutic equipment.
Select a room with no excessive noise, vibration, sunlight, high
humidity or water splashes.
When connecting the cables, make sure that each instrument is
turned off.
Only install the specified software in the electroencephalograph.
Otherwise the electroencephalograph may malfunction.
Make sure that there is no influence from a cellular phone.
Avoid locations where the wireless input unit may receive strong
electromagnetic interference such as radio or TV stations, cellular
phones or mobile two-way radios.
A sudden loss of power or extreme power surge can damage data.
To assure an uninterrupted power supply, use an uninterruptable
power supply (UPS).
Do not install the wireless input unit where it will be exposed to
water or chemical solutions. Avoid direct sprinkling, spray or moist
air from the nebulizer or humidifier. These cause malfunction and
shorten the life of the unit.
After installing the telemetry unit and access point, check that the
communication between the telemetry unit and access point is
correctly performed without any interference.
For external instrument connection and local area network connection, refer to
“General Requirements for Connecting Medical Electrical Systems” in this section.
NOTE
Do not place blankets or cloth over the access point.
Do not install the wireless input unit in dusty area.
Connect the power cable to an AC outlet which can supply enough
AC current to the access point. The access point cannot function
properly with low current.
For electroencephalograph installation, refer to the operator’s manual for the
electroencephalograph.