2. INSTALLATION/PREPARATION
2.6 Operator's Manual WEE-1000
Cable Connection
WARNING
When connecting the cables, make sure that all components of the
wireless input unit are turned off.
Connect only the specified instruments to the connectors or socket
marked with
, by following the specified procedure. Otherwise,
electrical leakage current may harm the patient and operator.
Connecting to a Local Area Network
When connecting the access point and electroencephalograph with a
local area network, connect the access point and
electroencephalograph so that the access point and
electroencephalograph are electrically separated from the local area
network according to the IEC 60601-1-1 “Medical electrical equipment
- Part 1-1: General requirements for safety - Collateral standard:
Safety requirements for medical electrical systems”. Failure to follow
this warning may cause electrical shock to the patient and operator.
Check that there is no damage on the surface of the network cable. If
it is damaged, it may cause electrical shock to the patient and
operator.
1. Connect the provided network cable to the PC connector on the left side panel
of the access point.
Connecting the Access point
and Electroencephalograph
2. Connect the other side of the network cable to the network connector on the PC
unit of the electroencephalograph.
PC connector