2. INSTALLATION/PREPARATION
Operator's Manual WEE-1000 2.7
Connect the provided power cord to the AC SOURCE socket on the right side panel
of the access point and plug the cord into a 3-prong AC outlet.
CAUTION
Only use the provided power cords. If another power cord is used, it
may cause electrical shock.
WARNING
For patient safety, equipotential grounding may be required. Consult
with a qualified biomedical engineer.
When more than one electrical instrument is used, there may be electrical potential
difference between the instruments. Potential difference between instruments may
cause current to flow to the patient connected to the instruments, resulting in
electrical shock (micro shock). Never use any medical equipment in patient
treatment without proper grounding.
Always perform equipotential grounding when required. It is often required in the
operating room, ICU room, CCU room, cardiac catheterization room and X-ray
room. Consult with a biomedical engineer to determine if it is required.
When Equipotential Grounding is Required
Connect the equipotential ground terminal on the right side panel of the access
point to the equipotential ground terminal on the wall with the ground lead.
Connecting the Power Cord
Connecting the Power
Cord
Equipotential Grounding