User Manual

2. INSTALLATION/PREPARATION
2.14 Operator's Manual WEE-1000
Connecting the Telemetry
unit to the Access point
with the Isolator
The telemetry unit can also be directly connected to the access point with an SC-
101A Isolator. DC power is supplied to the telemetry unit from the access point
through the isolator.
CAUTION
When connecting the isolator to the telemetry unit, make sure that
telemetry unit is turned off.
When using the isolator, do not remove the battery holder.
Otherwise, the telemetry unit may malfunction due to static energy.
When moving the patient, make sure that the cable connected
between the isolator and access point is disconnected. Otherwise,
the patient may fall over the cable, or the cable may get broken.
When connecting or disconnecting the connection cable to/from the
isolator, hold the connectors. If you hold the cable, the cable may
get broken.
1. Disconnect the connection cable from the isolator which connects the isolator
and access point.
2. Connect the cable from the isolator to the telemetry unit and secure the
connector with the two screws. Use the provided flat-blade screwdriver to
secure the connector.
3. Connect the connection cable to the access point and secure the connector with
the two screws.
4. Connect the AC power cord of the access point to the AC outlet.
5. Connect the other side of the connection cable to the isolator. DC power is
supplied to the telemetry unit and the telemetry unit is turned on. The LCD
display displays the version information, then displays “CL”.
To turn off the telemetry unit, disconnect the connection cable from the
isolator. When the battery is set in the battery holder, the telemetry unit
continues to operate on battery power.