2. INSTALLATION/PREPARATION
Operator's Manual WEE-1000 2.15
Power On/Off Procedure
CAUTION
When connecting the isolator to the telemetry unit, make sure that
telemetry unit is turned off.
When using the isolator, do not remove the battery holder.
Otherwise, the telemetry unit may malfunction due to static energy.
1. Before turning the power on, check the following items. If there is any damage
or the instrument is suspected to be faulty as a result of checking, attach an
“Unusable” or “Repair required” label to the instrument and contact your NK
distributor or representative.
Check items before turning the power on
Overview:
Telemetry unit, access point and isolator are not dirty, damaged or in contact
with liquid.
Power cord is not damaged.
No key on the telemetry unit is broken.
No electrode is dirty or damaged.
No electrode lead is frayed or damaged.
No connection cable is frayed or damaged.
Connection and Setting:
Telemetry unit
Fully charged lithium-ion battery, a new lithium battery or a new alkaline
battery is inserted.
Battery is inserted in the correct direction.
Battery holder is attached to the telemetry unit firmly.
Isolator is firmly connect to the telemetry unit.
Electrode junction box is properly connected to the telemetry unit.
Access point
Power cord and ground lead are properly connected to the access point.
Access point is properly connected to the electroencephalograph with the
network cable.
Environment
There is no obstruction between the telemetry unit and access point.
Bluetooth device or other wireless LAN device is not used near the wireless
input unit.
Power On Procedure