2. INSTALLATION/PREPARATION
2.16 Operator's Manual WEE-1000
Accessories:
Enough electrodes.
Enough EEG paste.
A second battery.
2. Turn the power on
Wireless communication:
1) Connect the AC power cord of the access point to the AC outlet.
2) Press the START/OK key on the telemetry unit.
Wired communication:
1) Connect the telemetry unit and access point with the isolator.
2) Connect the AC power cord of the access point to the AC outlet.
3. After turning the power on, check the following items.
Check items after turning the power on
Overview:
No fire, smoke or smell.
No electrical shock when touching the telemetry unit or access point.
Telemetry unit and access point are not too hot.
Telemetry unit and access point do not affect surrounding equipment.
Telemetry unit
No beep sounds.
No error code is displayed on the LCD screen.
All keys operate correctly.
Access point
The power lamp lights.
TX and RX lamps blinks
All operation indicators light.
Electroencephalograph
The communication between the telemetry unit and access point is
established with or without the isolator.
The message to replace the battery of the telemetry unit is not displayed.
There is no error message on the screen or malfunction.
The screen display is correct.
The time and date display is correct.
All programs operate correctly.
All settings (such as montage and amplifier settings) are correct.
Calibration waveform is properly displayed.
No noise on the calibration waveform.
Amplitude of the calibration waveform is correct.