User Manual

1. GENERAL
1.8 Operator's Manual WEE-1000
Battery
DANGER
Keep the battery away from fire. Do not heat the battery. Otherwise,
the battery explodes.
Do not immerse the battery in water or seawater. The battery heats up
and rusts and the battery liquid leaks out.
Never use a battery which is damaged, discolored or has leakage. A
damaged battery explodes if used. If the battery is damaged and the
battery liquid contacts the eyes or skin, wash immediately and
thoroughly with water and see your physician. Never rub your eyes,
otherwise you may lose your eyesight.
Never disassemble, modify or give impact to the battery. The battery
short-circuits and the battery liquid leaks out.
Never short-circuit the + and – terminals on the battery with a wire.
Do not leave the battery where patients can reach it. If a battery is
swallowed, see your physician immediately.
Do not expose the battery to direct sunlight or leave it in a high
temperature place, such as in a car or near heater . The lifetime of the
battery may be shortened, the performance of the battery may be
degraded and the battery liquid may leak out.
WARNING
Only charge the CGR-B/242 rechargeable lithium-ion battery with the
provided DE-158UA Battery Charger.
Charge the rechargeable lithium-ion battery at the surrounding
temperatures of 10 to 40°C (50 to 104°F). If the battery is charged
below 10°C or over 40°C, it may leak or heat up. This may damage the
battery.
CAUTION
Battery replacement should only be done by the operator. During
measurement, when replacing the battery, be careful not to touch the
patient.
Do not charge a deteriorated battery. Otherwise, the instrument cannot
operate on battery power.