3. EEG/PSG MEASUREMENT
Operator's Manual WEE-1000 3.11
5. Take the line passing over F7, Fz and F8 and attach an electrode at the mid-
point (F3) between F7 and Fz as shown in Fig. 4. Attach another electrode at
the mid-point (F4) between F8 and Fz.
6. Take the line passing over T5, Pz and T6 and attach an electrode at the mid-
point (P3) between T5 and PZ. Attach another electrode at the mid-point (P4)
between T6 and PZ.
Earlobe Electrodes
1. Apply a small amount of EEG paste to both earlobes. Gently clip on the
earlobe electrodes.
2. Secure the electrode lead to the neck with surgical tape to prevent electrode
lead movement.
CAUTION
Do not apply excessive pressure to the electrode because this may
cause direct skin-electrode contact (not through the paste), and
generate unnecessary polarization voltage.