3. EEG/PSG MEASUREMENT
3.16 Operator's Manual WEE-1000
Surgical tape
Wrist strap
Loop
Slack in the sensor lead
Pocket for the
PLM sensor
Periodic Limb Movements (PLM)
The PLM are measured from the limb or leg (heel) by
the disk electrode or PRO-Tech PLM sensor.
Body Position
The sleeping body position is measured by the body
position sensor attached on the clothing or surface of
the body. The Pro-Tech Sleep Position Indicator is
recommended for the measurement.
Take care to attach the sensor to the patient so that
the patient dose not become uncomfortable.
Sleep Position Sensor
Top-Head
This Side Out
--
Surgical tape
Sensor lead
Body position sensor
attached to the chest
respiration sensor strap
Snore
The snore is measured from the neck by the airway vibration. The Pro-Tech
Dynamic Snoring Microphone is recommended for the measurement.
Take care to attach the sensor because the carotid artery pulse waveforms are
measured as artifact if the sensor is attached around the carotid artery.
Surgical tape
Snoring sensor
Slack in the sensor lead
Electrode junction box
PLM Sensor
Pocket
PLM Strap
Lead Wire
Stress Loop
Medical Tape
Strain Relief
Rectangular
Loop