3. EEG/PSG MEASUREMENT
Operator's Manual WEE-1000 3.19
Connect the electrodes and sensors to the electrode junction box. Two types of
electrode junction box can be used.
JE-011A Electrode Junction Box
30 channels of EEG waveforms or 22 channels of EEG waveforms, 8 channels of
bipolar signals and 2 channels of DC input signals.
JE-012A Electrode Junction Box
For polysomnogram (PSG) measurement
JE-013A Electrode Junction Box
62 channels of EEG waveforms and 2 channel of DC input signals
NOTE
Do not bend or pull the lead or cables. This may damage them.
The following extra jacks can be used as bipolar jacks. To select
extra jacks or bipolar jacks, refer to the System Program.
- JE-011A Electrode Junction Box: X2 to X9
- JE-012A Electrode Junction Box: X17 to X24
When using the extra jack as the extra jack, only the (+) jack is
available. When using the extra jack as the bipolar jack the (+) and (-)
jacks are available.
For electrode combination (montage), refer to “Changing System Settings -
Programming Pattern” in Section 4 of the operator’s manual for each
electroencephalograph .
Connecting the Electrodes and Sensors to the Electrode Junction
Box