Operator’s Manual ZM-520PA/521PA/530PA/531PA 33
Changing System Setup Settings
PROTOCOL
Select the transmitting protocol. For differences between protocols, refer to the table below.
51: New protocol. A central monitor with ORG-9100A/9110A multiple patient receiver
whose software version 03-03 or later can receive this protocol.
41: Old protocol. A central monitor with ORG-9100A/9110A/9700A multiple patient receiver
can receive this protocol.
NOTE
When 51 is set, the receiving monitor must be able to receive protocol 51. Otherwise,
signals from the transmitter cannot be received.
Differences Between Protocols
Function Protocol 41 Protocol 51
Setting ECG MEASUREMENT to OFF on
the transmitter automatically turns off the
ECG measurement setting on the receiving
monitor
No (ECG measurement
must be turned off on
the receiving monitor)
Yes
Pause monitoring on the receiving monitor
from the transmitter
No Yes
Transmit “PATIENT CONFIRMED” message No Yes
Display battery level of the transmitter on the
receiving monitor
No Yes
Transmit SpO
2
messages
Some messages (refer
to the “Indication and
Message List” section)
All messages
Transmit RESP message No Yes
1. Press the
key to move the cursor to
“PROTOCOL”.
2. Press the Function key.
3. Press the
key to select “51” or “41”.
Setting
Cursor
Setting
Cursor