User's Manual

Operator’s Manual ZM-520PA/521PA/530PA/531PA 17
CAUTION
If the attachment site is dirty with blood
or bodily fluids, clean the attachment
site before attaching the probe. If there
is nail polish on the attachment site,
remove the polish. Otherwise, the
amount of transmitted light decreases,
and measured value may be incorrect
or measurement cannot be performed.
CAUTION
If the skin gets irritated or redness
appears on the skin from the probe,
change the attachment site or stop
using the probe. Take extreme care for
the patients with delicate skin.
CAUTION
Failure to follow these instructions may
cause cable discontinuity, short circuit,
skin burn on the patient from the probe
temperature increase due to the short
circuit of the probe cable, and incorrect
measurement data. If the probe is
broken, replace it with a new one.
• Do not immerse any part of
the probe cable other than the
disposable probe in chemical
solutions or water.
• Do not pull or bend the probe cable.
• Do not let caster feet run over the
probe cable.
CAUTION
When the probe is attached on an
appropriate site with sufficient
circulation and the error message
confirming the probe attachment
repeatedly appears, the probe may be
deteriorated. Replace it with a new
one.
Maintenance
CAUTION
This transmitter is not waterproof. If
detergent or liquid spills into the
transmitter, stop cleaning or
disinfecting it and contact your Nihon
Kohden representative. The transmitter
needs to be checked for safety and
function before use.
CAUTION
Do not immerse the electrode lead
connector in liquid.